Skip to content

The Lodestar DX system

A bench-top molecular analyser and single-use test strip. One test, six pathogens, 35 minutes.

LAMP in a closed tube at constant temperature.

The test strip holds six pathogen assays plus positive and negative controls. Both controls are read on every run.

Lodestar DX near-patient UTI analyser with test cartridge loaded by clinician

Five steps to a result.

  1. Urine sample in a specimen pot

    Collect a urine sample. Samples should be processed within 24 hours of collection.

  2. Urine being pipetted into the diluent tube

    Transfer 10 µL of urine into the supplied diluent tube and mix thoroughly.

  3. Diluted sample being pipetted into the test strip tubes

    Transfer 5 µL of the mixture into each of tubes 1 to 7 on the test strip. Tube 8 is the negative control and takes no sample.

  4. Test strip being loaded into the analyser

    Seal the tubes, mix, load the strip into the analyser and start the run. The run takes 35 minutes.

  5. Analyser display showing result indicators

    Read the display. Red indicates the target was detected at clinically relevant levels, green indicates it was not. Record the result: results are not stored on the analyser.

Six of the most common uropathogens, identified in one run.

The Lodestar human UTI test panel detects DNA of six uropathogens that account for the majority of community and hospital UTI presentations. See clinical performance data.

Lodestar DX six-pathogen panel: Escherichia coli, Enterococcus species, Staphylococcus saprophyticus, Proteus mirabilis, Pseudomonas aeruginosa, Klebsiella pneumoniae

Escherichia coli · Enterococcus species · Staphylococcus saprophyticus · Proteus mirabilis · Pseudomonas aeruginosa · Klebsiella pneumoniae

The panel does not include Group B Streptococcus. Cross-reactivity is documented in the IFU: the E. coli assay may also detect Shigella species, the Enterococcus assay may detect some strains of Staphylococcus aureus or S. saprophyticus, and the K. pneumoniae assay may detect some strains of K. variicola or K. oxytoca.

Built for clinical use.

Sample volume
10 µL urine
Time to result
35 minutes
Throughput
One test strip, one patient, per 35-minute cycle
Technology
Loop-mediated isothermal amplification (LAMP)
Targets
Six uropathogens plus positive and negative controls
Limit of detection
Between 10⁴ and 10⁵ cfu/mL across all six assays
Test strip storage
Supplied and stored cold chain; refer to the Instructions for Use
Connectivity
Can integrate with common EPR systems via EKF Link, subject to local IT configuration
Quality control
QC material available (QC-HUTI-10); monthly QC recommended, or on starting a new lot
Operators and setting
Trained healthcare professionals, in laboratory or near-patient settings
Regulatory
UKCA-marked. ISO 13485:2016 (BSI certificate MD 812652). GMDN 64635.

Specifications correct at time of publication. Refer to the Instructions for Use for full technical detail.

Installation, training and ongoing support.

Installation, operator training and ongoing support are provided by Syner-Med. Training is required before first use.

Change the way you manage UTIs.

Talk to Syner-Med about implementing the Lodestar UTI testing system in your clinic or lab.